INCISE® INTRAOCULAR LENS - Less Invasive Surgery, More Rapid Recovery

MICRO-INCISION ONE-PIECE HYDROPHILIC ACRYLIC IOL

INCISE® combines high quality optics with new 1.8 mm in the bag or 1.4 mm wound assisted technique implantation, enabling you to perform sub 2 mm MICS™ with ease:

  • Designed to minimise PCO with sharper 360° posterior barrier edge (≤ 5 microns radius)
  • Stiffer four-point fi xation haptic design for stability and centration in the capsular bag
  • Aspheric advanced optics designed to enhance visual quality
  • Four bridged haptics to minimise the effect of post-operative capsular bag contraction
  • Controlled unfolding facilitates precise positioning in the capsular bag and removal of viscoelastic

Product Information

MATERIAL

  • Unique advanced acrylic
  • 22 % water content
  • UV-blocker
  • Refractive Index: 1.47

DESIGN

  • Monofocal Aberration-Free Aspheric Optic
  • Sharper 360° posterior barrier edge (≤ 5 microns radius)
  • Angulation: 3° to 10° haptic angulation across the range
  • 4-point fi xation haptic design
  • Orientation features to indicate the anterior side
  • Optic diameter: 6.0 mm for a +20.0 D
  • Overall diameter: 11.0 mm: 0.0 D to +30.0 D

DIOPTER RANGE

From 0.0 D to +30.0 D: 0.0 D to +10.0 D in 1.0 D increments; +10.0 D to +30.0 D in 0.5 D increments

INJECTORS

  • Viscoject™ BIO 1.5 LP604361C (10/box); Recommended incision size: 1.4 mm WAT or 1.8 mm in the bag
  • Viscoject™ BIO 1.8 LP604350C (10/box); Recommended incision size: 1.8 mm WAT

CONSTANTS*

  • Immersion A-Scan or IOLMaster: A-Constant SRK/T: 118.9; ACD: 5.51; Surgeon Factor: 1.75; Haigis Constant: a0: 1.35 / a1: 0.40 / a2: 0.10
  • Applanation A-Scan: A-Constant: 118.4; ACD: 5.37; Surgeon factor: 1.62

* Constants are estimates only (source: ULIB Optimized IOL Constant, www.augenklinik.uni-wuerzburg.de/ulib/c1.htm) It is recommended that each surgeon develops their own values. Latest update: June 2017