EyeCee® ONE INTRAOCULAR LENS

ONE-PIECE HYDROPHOBIC ACRYLIC IOL

Fully Preloaded Hydrophobic IOL

  • Easy and convenient handling of the preloaded system with usage of low molecular weight OVD (please refer to the loading guide)
  • 2.4 mm in the bag
  • Single use injector

Quality of Vision

  • Asperitic optic edge for reduced post-operative glare
  • Negative aspheric optic for enhanced contrast sensitivity
  • 90° anchor-wing haptics to improve capsular bag fixation and prevent lens rotation
  • Blast finishing haptic design
  • 360° sharp edge to reduce risk of PCO
  • Blue light filtering

Indications & approvals may vary by country. Please contact your local representative for further information.

Product Information

MATERIAL

  • Hydrophobic Acrylic
  • UV and blue light blocker
  • Refractive index: 1.52

DESIGN

  • Monofocal Negative Aspheric Optic
  • Modifi ed C-loop
  • 360° posterior square edge
  • Optic diameter: 6.0 mm
  • Overall diameter: 13.0 mm

DIOPTER RANGE

  • Preloaded from +11.0 D to +30.0 D: +11.0 D to +27.0 D in 0.5 D increments; +27.0 D to +30.0 D in 1.0 D increments; Recommended incision size: 2.4 mm in the bag (please refer to the DFU)
  • Non preloaded from +1.0 D to +30.0 D (+1.0 D to +10.0 D in 1.0 D increments; +10.0 D to +27.0 D in 0.5 D increments; +27.0 D to +30.0 D in 1.0 D increments)

INJECTOR

  • MDJ 2.0 - 2.2 MDJ20-22 (1/box); Recommended incision size: 2.2 mm WAT
  • MDJ LOADINJECT® 2.2 MDJLOAD22 (1/box); Recommended incision size: 2.2 mm WAT

CONSTANTS*

  • Immersion A-Scan and IOL Master: A-Constant SRK/T: 119.7; ACD: 6.0; Surgeon Factor: 2.13; Haigis Constant: a0: 1.675 / a1: 0.40 / a2: 0.10
  • Applanation A-Scan: A-Constant: 119.1; ACD: 5.70; Surgeon Factor: 1.73